The Guardian 6100 is designed for fast, easy, production testing of single phase electronic medical devices and instruments. Without changing the connection to the device, the tester is able to perform five essential electrical safety tests in one unit, including AC Hipot, DC Hipot, Insulation Resistance, Line Leakage Current and Ground Bond measurements. Performing all these electrical safety tests in one box increases production by reducing test time and eliminating multiple test stations.
AC Hipot Test: Performs AC dielectric testing (hipot) over the voltage range from 50V to 5000V AC RMS. Leakage current detection is programmable from 1ìA to 40mA, making the 6100 ideal for testing medical products and instruments having a wide range of leakage currents.
DC Hipot Test: DC dielectric testing is possible from 50V to 6000V DC with leakage current detection down to 0.1ìA. The maximum total output cutrrent of 20mA allows quick charging of capacitive devices and products.
Insulation Resistance: The insulation resistance test (similar to a DC hipot) calculates and displays a product¿s insulation resistance value in ohms. This resistance can be measured over the range of 100kÙ to 50GÙ with test voltages programmable from 50V to 1000V DC.
Ground Bond: The ground bond test, sometimes referred to as a high current continuity test, can be programmed from 1A to 30A AC for verifying the integrity of a product¿s ground system. Resistance measurements are displayed and a high limit is programmable between 0.1mÙ and 510mÙ.
Line Leakage Current: Line leakage tests (Earth, Patient and Enclosure) can be performed directly on the Guardian 6100 in eight possible configurations including: normal operating conditions, reverse line, or a single fault normal and single fault reverse mode with ground set on or off. Five different human body circuit models can be selected in accordance with UL, IEC, and other standards.Leakage current limits are programmable from 0.1ìA to 9.999mA.
Options included:
Item | Description | |
---|---|---|
Chroma A190314 | Chroma A190314 Quadtech G32 1000VA Isolation Transformer. | |
QuadTech G31 | Chroma Quadtech G31 500VA Isolation Transformer. |
Item | Description | Price | |
---|---|---|---|
QuadTech Guardian 6000 | Quad Tech Guardian 6000 Production Safety Analyzer AC/DC Hipot, IR, Ground Bond. | $1995.00 | |
QuadTech Option 6000-02 | Chroma Quadtech Guardian A190303 Option 6000-02 scanner card for guardian 6000 Plus. (Output cables available at additional cost). Eight Channels. 3 High Voltage, 5 Ground Continuity. | $295.00 |
QuadTech Guardian 6100