The Agilent 85025A detector is designed specifically for operation with the 8757 series scalar network analyzer and is not compatible with the 8756 or the 8755. The 85025A detects either a modulated (ac) or an unmodulated (dc) microwave signal.
85205A detector specifications when used with an 8757 series scalar network analyzer:
- Frequency range: 0.01 to 18 GHz
- Dynamic range
- AC mode: +16 to -55 dBm
- DC mode: +16 to -50 dBm
- Nominal impedance: 50 Ù
- Maximum input power: +20 dBm (100 mW), ±10 VDC
- Return loss
- 10 MHz to 40 MHz: 10 dB
- 40 MHz to 4 GHz: 20 dB
- 4 GHz to 18 GHz: 17 dB
- Frequency response in DC mode, input power -10 dB
- 10 MHz to 40 MHz: +0.25 dB/ -0.75 dB
- 40 MHz to 18 GHz: ±0.5 dB
- Cable length: 1.22 m (48 inches)
- Recession of the male center conductor from the reference plane
- 85025A: 0.207 to 0.210 inches
- 85025A Option 001: 0.000 to 0.003 inches
Please contact our sales department for more information.