To evaluate components at currents up to 125 A the 3265B DC Bias Units are used with either the Wayne Kerr 3255B series of Inductance Analyzers or the 3260B Precision Magnetics Analyzer. When one 3265B/25A or 3265BQ/25A DC Bias Unit is connected to an instrument up to 25 A of DC bias current can be set in steps of 25 mA with one unit. Additional DC Bias Units can be added such that with five units connected in parallel it is possible to set DC bias currents up to a maximum of 125 A DC.
The 3255BL may only be used with the 3265B/5A and 3265B/10A, 5A and 10A DC bias units. Multiple instruments may be connected in parallel to give a maximum DC bias current of up to 50A which may be set in steps of 25 mA.
The instruments have a number of safety and protection features including a safety interlock system to protect the user against back EMFs. They are also fully protected against over temperature, excess voltage drop and sense lead failure.
With the addition of the 1009 DC Bias Fixture DC bias currents up to 50 A can be applied to an SMD inductor during component test in order to evaluate the devices thoroughly at the operational bias currents.
The fixture operates with either one or two 3265B DC bias units and a 3260B Precision Magnetics Analyzer. The optional 1009 high current lead set will be required if two 3265B/25As are used.
Four front panel mounted BNC connectors and two captive high current cables ensure simplicity and ease of use with a 3265B.
Interchangeable component test carriers ensure that the 1009 DC Bias Fixture may be used with a wide variety of devices. Blank carriers are available which enable device specific test fixtures to be developed or alternatively a carrier design and manufacturing service is available.
Measurement Function | In impedance mode: L, Z, R, Q, D. Not applicable to Rdc, or transformer measurements | |||||
Frequency Range | 3255BL 20 Hz to 200 kHz (not compatible with 3265B/25A and 3265BQ/25A) 3255B 20 Hz to 500 kHz 3255BQ 20 Hz to 1 MHz 3260B 20 Hz to 3 MHz (3265BQ/25A) | |||||
Compliance Voltage | Maximum compliance votage (<12 kHz) 11 V DC at 0.25 V AC drive level. 10 V compliance at 1 V AC drive level. For f>12 kHz deduct 0.5 V. | |||||
Basic Accuracyy | ±1%, Varies with measurement speed, frequency and options chosen. | |||||
Users Manual (QR Link)
Power Cord Coax Cable, RG-58, 8" (Qty 4) Control Link Cable Fuse
Contact InfoPlease let us know how we can contact you.Additional InfoIn-Stock Notification Request SentYour request to be notified when this product is back in stock has been sent to our sales department. We will notify you as soon as this item is available for purchase.Thank you for allowing us the opportunity to assist with your test equipment needs. We hope to earn your business. Error Sending RequestWe are very sorry but an error on the server is preventing your request from being processed.Please contact us via email or toll free at (800) 673-4102 and we will be happy to assist you with this request. We would also greatly appreciate it if you could provide us with this error code: 999 Unable to Process RequestYour offer was received but we are unable to process it for the following reason:??? Please correct the issue indicated and resubmit or contact us via email or toll free at (800) 673-4102 and we will be happy to assist you with this request. OfferAccuSource Electronics is always happy to consider reasonable offers on our test instruments. Please enter your offer below for consideration.Wayne Kerr 3265B
$
.00
Contact InfoRequirementsOur normal service includes 90 day warranty, all standard accessories and manuals, and restoration to "nearly new" condition. Discounts may be offered for reduced service levels. Please let us know which of these you would be willing to consider for the price discount.Reduced Warranty Period: Do you require manuals and accessories for the instrument? Will you accept an instrument with minor cosmetic and physical flaws if it operates properly? Are you willing to prepay your order by check or wire transfer to avoid credit card fees? Additional InfoEquipment Offer SentYour offer has been sent to our sales department for review. You should expect a response within one business day or less.Thank you for allowing us the opportunity to assist with your test equipment needs. We hope to earn your business. Error Sending RequestWe are very sorry but an error on the server is preventing your request from being processed.Please contact us via email or toll free at (800) 673-4102 and we will be happy to assist you with this request. We would also greatly appreciate it if you could provide us with this error code: 999 Unable to Process OfferYour offer was received but we are unable to process it for the following reason:??? Please correct the issue indicated and resubmit or contact us via email or toll free at (800) 673-4102 and we will be happy to assist you with this request. |