AC Hipot
| Voltage Range | 0.100KV to 5.000KV |
| Voltage Setting | 5 Volt Increments |
| Voltage Regulation (Line & Load) | 1% + 5V |
| Voltage Accuracy | 1% of reading + 10V |
| Current Sourcing | 30 to 40mA (above 500V, 180sec. max.) 0 to 29.99mA (above 500V, continuous) 0 to 10mA (below 500V, continuous) |
| Current Limit Range | 0.10 to 40mA, 0.02mA/step |
| Current Accuracy | 1.5% of reading + 50µA |
Continuity Measurement
| Test Current | 100mA |
| Resistance Range | 0.001 to 1.500Ù |
| Resistance Accuracy | ±0.075Ù |
General
| Test Database | 10 test sequences with up to 16 steps per test sequence |
| Data Lockout | Prevents database edit while in lockout mode |
| Arc Detect | 40 level (1 to 40mA) |
| Serial Port | 9600 Baud, fully bi-directional RS232 interface with PC cable included |
| GPIB | Option GP-60, IEEE-488 interface |
| PLC Interface | 9-pin D-sub connector with outputs of Testing, Pass, Fail and inputs of Start, Reset. Output isolation is 400V with 130mA continuous load current |
| Scanner Interface | Control port for V16 and V17 HV Scanners |
| Timer Accuracy | 0.25% ± 50mS |
| Power Source | AC 100V, 120V, 220V, 230V at ± 10% 50/60Hz |
| Operating Environment | Indoor use, altitude up to 6500ft (2000m). Installation Category II. Pollution Degree 2 |
| Operating Temperature | 0°C to 40°C, 80% RH max |
| Storage Temperature | -10°C to 70°C, 70% RH max |
| Dimensions | 17.6"L x 13"W x 5.9"H) 446L x 330W x 149H mm |
| Weight | 30 lbs (14kg) Net, 34 lbs (16 kg) Ship |
| Accessories | NIST certification, TL-20-60 Alligator Test Leads, serial interface cable, instruction manual and power cord |
| Accuracy | Stated specifications apply for one year at 23°C ± 5°C |
| Calibration | ANSI/NCSL Z540 Certificate of calibration, with data, traceable to NIST provided |
| Compliance | CE Mark Certified to EN61010 and EMC directive EN50081-1 & EN50082-2 |
Vitrek V60