The 6330 is Slaughter's most advanced multi-function
electrical safety tester with 6-in-1 test capability.
This flagship model is designed to make advanced
testing applications simple with the most feature rich
menu system on the market. With minimal set-up
time, you can perform AC Hipot, DC Hipot, Insulation
Resistance, Ground Bond, Leakage Current and
Functional Run tests on a variety of DUT’s.
The 6330 maximizes production line throughput far
beyond a single function tester, while taking up less
space. Use the 6330 from the front panel or automate
your testing via the RS-232 interface.
- Hipot:
- 3.5 kV @ 30 mA AC
- 4.0 kV @ 5 mA DC
- Insulation Resistance:
- Range: 100 - 1000 VDC
- Accuracy: ± (2% of reading + 5 V)
- Ground Bond:
- Output Voltage Range: 6 VAC fixed
- Output Frequency Range: 50/60 Hz, user selectable
- Output Current Range: 3.0 - 30.0 AAC
- Current Display Range: 0.0 - 30.0 A
- Run Test:
- DUT Power
- Voltage: 0 - 277 VAC Single Phase Unbalanced
- Current: 30 AAC max continuous
- Leakage Current Test:
- DUT Power
- Voltage: 0 - 277 VAC Single Phase Unbalanced
- Current: 30 AAC max continuous
Operation Manual (QR Link)
Service Manual (QR Link)
Power Cord
High Voltage Cable
Interlock Connector (installed)
Adapter Box
High Current Output Lead 10'
High Current Return Lead 10'
High Current Output Lead 6'
High Current Return Lead 6'
LLT DUT Input Cable Line
LLT DUT Input Cable Neutral
LLT DUT HV Output Cable Line
LLT DUT HV Output Cable Neutral
RS232 Cable