HP 432A/B Power Meters
High accuracy—no thermoelectric error: high accuracy over a wide temperature range is featured on the HP 432 Power Meters. By measuring the output voltage of the thermistor bridges, and computing the corresponding power, even higher accuracy of ±0.2% ±0.5 uW can be obtained.
Accuracy can be maintained on even the most sensitive range because the error due to thermoelectric effect is reduced to a negligible level.
Calibrated mounts: each thermistor mount is furnished with data stating the Calibration Factor* and Effective Efficiency* at various frequencies across the operating range. For easy and accurate power measurements, the front panel of the HP 432 contains a calibration factor control, calibrated in 1% steps from 88% to 100%, that compensates for losses in the mount and eliminates the need for calculation.
Instrument type:automatic, self-balancing power meter for use with temperature-compensated thermistor sensor.
Item | Description | |
---|---|---|
Agilent / HP 478A | Agilent HP 478A Coaxial Thermistor mount. 10 MHz to 10 GHz. | |
HP / Agilent 8477A | Agilent HP 8477A Power Meter Calibrator for 432A | |
HP / Agilent 8478B | Agilent HP 8478B - Coaxial Thermistor Mount. 10 MHz to 18 GHz. Type N connector. | |
HP / Agilent 8478B | Agilent HP 8478B - Coaxial Thermistor Mount. 10 MHz to 18 GHz. Option 011 APC-7 connector. Includes APC-7 to type N adapter. |
HP / Agilent 432B